• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 STATLOCK LUER RETAINER, SELECT PADS 8IN LUER LOCK EXTENSION SET 25 BOX; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 STATLOCK LUER RETAINER, SELECT PADS 8IN LUER LOCK EXTENSION SET 25 BOX; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Defective Component (2292)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 01/24/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.
 
Event or Problem Description
It was reported by customer that the product is cracked.It appears to be the extension tubing itself which would be different than the statlock devices which would go over the tubing and hub.Leaking stat-lock arterial line device.Newly inserted arterial line stat-lock device noted to have blood backing up and leaking on pillow.Upon switching to new line, noted crack in tubing where blood was leaking from.Product issue with cracking causing blood to back up into tubing, inaccurate blood pressure, switched to new stat-lock with no issues.It was reported this occurred with two devices.One device was used on a patient and one was not used on a patient.This report addresses the device that was used on a patient.
 
Event or Problem Description
It was reported by customer that the product is cracked.It appears to be the extension tubing itself which would be different than the statlock devices which would go over the tubing and hub.Leaking stat-lock arterial line device.Newly inserted arterial line stat-lock device noted to have blood backing up and leaking on pillow.Upon switching to new line, noted crack in tubing where blood was leaking from.Product issue with cracking causing blood to back up into tubing, inaccurate blood pressure, switched to new stat-lock with no issues.It was reported this occurred with two devices.One device was used on a patient and one was not used on a patient.This report addresses the device that was used on a patient.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a leak in the extension set tubing was inconclusive due to the sample condition.Two photographs of the arterial statlock kit (art0420) were provided.The provided photographs showed the product type that was included in the affected medline ynhh arterial line insertion bundles.These photographs were not of the affected products and no damage was apparent in the returned photographs.There was reference in the email attachments to other photographs of the affected product.However, these photographs were not attached to the complaint file and therefore could not be evaluated at this time.In addition, physical samples were sent to medline and therefore, bd does not have access to these samples for evaluation.This complaint will be recorded for future trending and monitoring purposes.
 
Event or Problem Description
It was reported by customer that the product is cracked.It appears to be the extension tubing itself which would be different than the statlock devices which would go over the tubing and hub.Leaking stat-lock arterial line device.Newly inserted arterial line stat-lock device noted to have blood backing up and leaking on pillow.Upon switching to new line, noted crack in tubing where blood was leaking from.Product issue with cracking causing blood to back up into tubing, inaccurate blood pressure, switched to new stat-lock with no issues.It was reported this occurred with two devices.One device was used on a patient and one was not used on a patient.This report addresses the device that was used on a patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK LUER RETAINER, SELECT PADS 8IN LUER LOCK EXTENSION SET 25 BOX
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DAVOL SURGICAL INNOVATIONS -9616067
ave. roberto fierro #6408
parque industrial aeropuerto
cd. juarez, chih s.a. de c.v. 32690
MX   32690
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18822017
Report Number3006260740-2024-00877
Device Sequence Number4579424
Product Code FPA
UDI-Device Identifier00801741081354
UDI-Public(01)00801741081354
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K980994
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Study,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberART0420
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2024
Supplement Date Received by Manufacturer03/04/2024
05/16/2024
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date03/15/2024
05/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-