Catalog Number CEB231010A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Sepsis (2067)
Date of Event 07/24/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The following publisher was reviewed by gore: title: outcome of treatment using excluder iliac branch endoprosthesis (ibe) in our site source: the journal of japanese college of angiology 2023: 63(suppl.) p.S197.8 evar cases (age 83 ± 4.5 years, 5 cases of male) using gore® excluder® iliac branch endoprostheses (ibe) for common iliac artery aneurysm that were performed from february 2019 to january 2022 at our hospital were evaluated.6 of these cases were with abdominal aortic aneurysm.4 of these cases were with internal iliac artery aneurysm.Technical success was achieved in all cases, and contralateral internal iliac artery embolization was required in 6 cases.The average postoperative observation period was 801 ± 414 days, and although one patient had open conversion due to infection and one patient had graft occlusion in internal iliac artery, there was no aneurysm enlargement (5 mm or more), and no aneurysm-related death or rupture.Outcome using ibe were good.Although there are anatomical limitations, the use of ibe is likely to continue to increase.Additional information was provided for the infection and open conversion case as in below.On (b)(6) 2021, the patient underwent endovascular treatment of an abdominal aortic aneurysm and bilateral iliac artery aneurysms using gore® excluder® aaa endoprostheses and gore® excluder® iliac branch endoprostheses.When inserting the internal iliac component into a 12 fr gore® dryseal flex introducer sheath, the catheter was difficult to be inserted.However, the catheter was able to be inserted once a pull-through was released.The patient was tolerated the procedure.On unknown date, infection was observed.On (b)(6) 2021, the open conversion was performed.The patient was taking immunosuppressant (anti-rheumatic medicine), and it was suspected that this may have been a risk.
 
Additional Manufacturer Narrative
H.6: investigation findings code c19: a review of manufacturing records verified that the lot involved in this complaint met all pre-release specifications.H.6.: investigation conclusions: code d12 and d1001: it should be noted the gore® excluder® iliac branch endoprosthesis instructions for use (ifu) states ¿ the safety and effectives have not been evaluated in the following patient populations: patients with active systemic infections.¿ according to the gore® excluder® iliac branch endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to: infection (e.G., aneurysm, device or access sites) as gore was unable to determine which device/device component is involved in this reportable adverse event, the following additional devices will be identified in this report: device name: gore® excluder® iliac branch endoprosthesis.Lot # 22969375.Catalog # hgb161407a.Udi # (b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Attachment of literature article.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
kathryn irwin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key18822583
Report Number3013164176-2024-01998
Device Sequence Number20782508
Product Code MIH
UDI-Device Identifier00733132635283
UDI-Public00733132635283
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Literature
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/13/2023
Device Catalogue NumberCEB231010A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2024
Supplement Date Received by Manufacturer03/11/2024
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date03/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-