MAUDE Adverse Event Report: GRIFOLS DIAGNOSTIC SOLUTIONS INC. PROCLEIX ULTRIO ELITE ASSAY; HIV-1/HIV-2/HCV/HBV DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GRIFOLS DIAGNOSTIC SOLUTIONS INC. PROCLEIX ULTRIO ELITE ASSAY; HIV-1/HIV-2/HCV/HBV DEVICE Back to Search Results
Catalog Number 731774
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Customer (b)(6) university (saudi arabia) reported on 05feb2024 initial ultrio elite nonreactive results for donor sample using ln 706570.Serology results were reactive for hbsag and anti-hbc.Discriminatory dhbv tests were reactive.Repeat ue test using same tube was reactive.Donor admitted to having hbv.Donation was not used for transfusion.No remaining sample was available for investigation.Summary of testing: (b)(6) 2024.Sample (b)(6).Results: ultrio elite nonreactive s/co 0.12.Sample (b)(6).(serology tube from the same donor) results: hbsag reactive (3 replicates).Anti-hbc reactive (3 replicates).(b)(6) 2024.Sample (b)(6).(serology tube from the same donor) results: ultrio elite reactive s/co 12.34.Dhbv reactive s/co 23.49.In addition, the customer indicated that pcr testing took place in (b)(6) 2023.In this particular case, nat and serology were carried out on (b)(6) 2024.Since the conventional practice involves conducting pcr as a confirmatory test following nat and serology, it is highly probable that the pcr tested a sample obtained from a prior draw.Consequently, the pcr outcome from december 2023 will not be considered in assessing the performance of the ultrio elite assay.Risk analysis: severity: critical.A single blood screening false negative (fn) result can cause multiple infections.The severity of a blood screening false negative is critical.Probability: remote.The probability of a false negative result is remote.False negative results do occur but are rare.Overall risk: acceptable.Gdss-rsk-000044, panther assay specific risk analysis for bloodbank assays (rsk-00854) and product safety risk management procedure, 04-03-12-sop were reviewed to assess risk.New risk to patient safety was not identified.There were no related nonconformances noted in the manufacturing record or as part of a search of the nonconformance records for this lot number.Product impact: no impact to product was identified.The root cause of the initial nonreactive was not determined due to lack of sample.Review of the qc release data and a previous events search in the grifols complaint system indicate that the assay is working as designed.Customer impact: the impact to the customer was questioning the ultrio elite results.Donation was not used for transfusion.Grifols risk assessment is based on severity of a harm or impact to human health and the probability of occurrence of harm or impact to human health.Based on the severity rating of ·critical· and probability rating of ·remote·, the overall risk posed by a false negative ultrio elite result is determined to be acceptable.Investigational testing is in progress.Follow-up information for this report will be provided when available.
 
Event or Problem Description
Customer king abdulaziz university (saudi arabia) reported on 05feb2024 initial ultrio elite nonreactive results for donor sample using ln 706570.Serology results were reactive for hbsag and anti-hbc.Discriminatory dhbv tests were reactive.Repeat ue test using same tube was reactive.Donor admitted to having hbv.Donation was not used for transfusion.No remaining sample was available for investigation.Summary of testing: (b)(6) 2024.Sample (b)(6) results: ultrio elite nonreactive s/co 0.12.Sample (b)(6) (serology tube from the same donor) results: hbsag reactive (3 replicates) anti-hbc reactive (3 replicates).22jan2024.Sample 1109634 (serology tube from the same donor) results: ultrio elite reactive s/co 12.34.Dhbv reactive s/co 23.49.In addition, the customer indicated that pcr testing took place in (b)(6) 2023.In this particular case, nat and serology were carried out on (b)(6) 2024.Since the conventional practice involves conducting pcr as a confirmatory test following nat and serology, it is highly probable that the pcr tested a sample obtained from a prior draw.Consequently, the pcr outcome from (b)(6) 2023 will not be considered in assessing the performance of the ultrio elite assay.Risk analysis severity: critical.A single blood screening false negative (fn) result can cause multiple infections.The severity of a blood screening false negative is critical.Probability: remote.The probability of a false negative result is remote.False negative results do occur but are rare.Overall risk: acceptable.Gdss-rsk-000044, panther assay specific risk analysis for bloodbank assays (rsk-00854) and product safety risk management procedure, 04-03-12-sop were reviewed to assess risk.New risk to patient safety was not identified.There were no related nonconformances noted in the manufacturing record or as part of a search of the nonconformance records for this lot number.Product impact no impact to product was identified.The root cause of the initial nonreactive was not determined due to lack of sample.Review of the qc release data and a previous events search in the grifols complaint system indicate that the assay is working as designed.Customer impact the impact to the customer was questioning the ultrio elite results.Donation was not used for transfusion.Grifols risk assessment is based on severity of a harm or impact to human health and the probability of occurrence of harm or impact to human health.Based on the severity rating of ·critical· and probability rating of ·remote·, the overall risk posed by a false negative ultrio elite result is determined to be acceptable.Investigational testing is in progress.Follow-up information for this report will be provided when available.23may2024: no further info is available.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLEIX ULTRIO ELITE ASSAY
Common Device Name
HIV-1/HIV-2/HCV/HBV DEVICE
Manufacturer (Section D)
GRIFOLS DIAGNOSTIC SOLUTIONS INC.
4560 horton street
emeryville CA 94608
Manufacturer Contact
jin shi
10808 willow court
san diego, CA 92127
8582020806
MDR Report Key18822857
Report Number2032600-2024-00001
Device Sequence Number4462710
Product Code QHO
UDI-Device Identifier00859882007672
UDI-Public00859882007672
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
BL125652
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/15/2025
Device Catalogue Number731774
Device Lot Number706570
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/01/2024
Supplement Report FDA Received Date05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-