| Catalog Number |
0030121 |
| Medical Device Problem Codes |
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Unspecified Infection (1930); Inadequate Osseointegration (2646)
|
| Date of Event |
02/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
|
| |
|
Event or Problem Description
|
|
During a surgical procedure, the #5 flexible reamer broke while being used on a patient's elbow.Both pieces of the reamer were retrieved, and an x-ray was taken, which showed no abnormalities.The original intended procedure was the removal of hardware from the right elbow, insertion of a cement antibiotic spacer, irrigation and debridement of the right elbow, and bursectomy of the olecranon bursa.The user encountered problems related to device failure and malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: device code, clinical signs code: the reported event was confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: a visual examination of the device reveals wear, shearing, and divots in the poly bushing of the ulnar stem.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.The investigation determined that patient factors were the root cause, with the healthcare provider performing the revision surgery indicating that the patient's weight caused excessive mechanical loading on the implant.The revision surgery targeted not the metal portion of the ulnar stem, but rather the poly bushing, which was confirmed to be damaged.Without x-rays or additional information provided by the healthcare provider, the loosening of the metal stem cannot be confirmed.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
During a surgical procedure, the #5 flexible reamer broke while being used on a patient's elbow.Both pieces of the reamer were retrieved, and an x-ray was taken, which showed no abnormalities.The original intended procedure was the removal of hardware from the right elbow, insertion of a cement antibiotic spacer, irrigation and debridement of the right elbow, and bursectomy of the olecranon bursa.The user encountered problems related to device failure and malfunction.
|
| |
|
Search Alerts/Recalls
|