B. BRAUN MEDICAL INC. INTROCAN SAFETY®; CATHETER,INTRAVASCULAR,THE Back to Search Results
Catalog Number 4251644-020
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2024
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: it was reported that the introcan safety pur 20g, 1.1x32mm-us while being used on a patient were "bendy" during insertion on patient and shattered during use.The emergency department has reported that the 20-gauge introscan safety iv catheter lot number 0061864777 has been kinking in patients during the catheter insertion and occasionally shredding when pulling the needle out of the catheter.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number: (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Fifty-nine (59) unused samples were provided for investigation.The provided 59 samples were visually evaluated, and no product deviation was noted.The defect is not confirmed.Retained units were evaluated and passed the internal testing.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY®
Common Device Name
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key18823286
Report Number2523676-2024-00237
Device Sequence Number19594816
Product Code FOZ
UDI-Device Identifier04046964398971
UDI-Public(01)04046964398971
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K020785
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number4251644-020
Device Lot Number0061864777
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/02/2024
Supplement Report FDA Received Date05/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-