• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLEARCUT S INCISIONAL INSTRUMENTS (WITH SHARPS INJURY PROTECTION); KNIFE, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLEARCUT S INCISIONAL INSTRUMENTS (WITH SHARPS INJURY PROTECTION); KNIFE, OPHTHALMIC Back to Search Results
Catalog Number 8065771540
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/06/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during surgery, an ophthalmic blades were dull.Procedure details and patient impact have not been provided.
 
Additional Manufacturer Narrative
Nine unopened knives, in a protector tray, in a blister were received for the report of dull blades.Samples were visually inspected and found to be conforming.Functional penetration testing was performed and all nine sample were found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The returned unopened samples were found to be conforming, therefore dull blades as described in the complaint was not confirmed.However since the actual complaint samples were not returned, a root cause cannot be determined for the complaint as described by the customer.The unopened samples were found conforming; however because the actual complaint samples were not returned the exact root cause for this complaint is unknown, specific action with regards to this complaint cannot be taken.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARCUT S INCISIONAL INSTRUMENTS (WITH SHARPS INJURY PROTECTION)
Common Device Name
KNIFE, OPHTHALMIC
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18825039
Report Number2523835-2024-00219
Device Sequence Number7749226
Product Code HNN
UDI-Device Identifier00380657715404
UDI-Public00380657715404
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K110166
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065771540
Device Lot Number168NMF
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/21/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/06/2024
Supplement Date Received by Manufacturer05/07/2024
Initial Report FDA Received Date03/04/2024
Supplement Report FDA Received Date06/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1