• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. SECURE-C; SECURE-C ENDPLATE ASSEMBLY, 13X14, 0 DEGREE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. SECURE-C; SECURE-C ENDPLATE ASSEMBLY, 13X14, 0 DEGREE Back to Search Results
Model Number 714.200S
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being resubmitted for an incident that occurred earlier.A revision surgery was performed approximately one year after implantation to remove a secure-c endplate assembly and core from the c6-7 level.The secure-c implant is comprised of two components, secure-c endplate assembly (714.200s) and secure-c core (414.207s}.The fluoroscopic image, taken during the revision surgery, showed parallel spacing between the implant endplates indicating the potential presen ce of the core within the confines of the implant.Examination showed no evidence of structural defects other than minor scratches and gouge marks that are consistent with the use of surgical instruments during removal.The exact reason for the revision was not provided; however, it was stated that the implant was not broken prior to removal.No determinations could be made as to the reason for the removal.
 
Event or Problem Description
It was reported that a revision surgery was performed to remove a secure-c endplate assembly and core for unknown reasons.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECURE-C
Common Device Name
SECURE-C ENDPLATE ASSEMBLY, 13X14, 0 DEGREE
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key18825186
Report Number3004142400-2018-00020
Device Sequence Number3076010
Product Code MJO
UDI-Device Identifier00889095179095
UDI-Public00889095179095
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P100003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number714.200S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2018
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age35 YR
Patient SexFemale
Patient RaceWhite
-
-