This report is being resubmitted for an incident that occurred earlier.A revision surgery was performed approximately one year after implantation to remove a secure-c endplate assembly and core from the c6-7 level.The secure-c implant is comprised of two components, secure-c endplate assembly (714.200s) and secure-c core (414.207s}.The fluoroscopic image, taken during the revision surgery, showed parallel spacing between the implant endplates indicating the potential presen ce of the core within the confines of the implant.Examination showed no evidence of structural defects other than minor scratches and gouge marks that are consistent with the use of surgical instruments during removal.The exact reason for the revision was not provided; however, it was stated that the implant was not broken prior to removal.No determinations could be made as to the reason for the removal.
|