ANGIODYNAMICS, INC. VENACURE EVLT; VENACURE 400 MICRON PERFORATOR AND ACCESSORY VEIN ABLATION KIT Back to Search Results
Catalog Number H787EVLTPVAKUS5
Medical Device Problem Codes Fracture (1260); Device Damaged Prior to Use (2284); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A medical assistant reported an issue with a pvak -- 400 micron perforator and accessory vein ablation kit.It was reported that the kit contained only blue coating at the tip, without an internal, clear fiber.This was discovered while unpacking the device; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
The reported defective device has been received by the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Additional Manufacturer Narrative
The customer's reported complaint description of distal portion of fiber tip was fractured and detached was confirmed based on visual inspection of the returned fiber complaint sample.The distal portion of the fiber tip was broken off and not returned.The likely root cause of the fiber fracture is handling damage but when and how this occurred cannot be definitively determined.The fiber tip is packaged such that the tip of the fiber is located under the coil wrap.A potential contributing factor for the fiber fracture is the coil wrap and/or the process of end user removing the coil wrap from the fiber during the unpacking process.Manufacturing personnel 100% visually inspect devices during the packaging process.This type of fiber fracture/detached damage would be noticed prior to shipment.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (16601420-01) which is supplied to the end user with the reported catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".The user manual (man/31/0075 us), which is supplied to the end user with this unit, states "before using a fiber, check it carefully for any signs of damage during storage or transit.Protective caps should be in place over sma connectors.Do not use if there is any sign of damage." a review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference pr(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENACURE EVLT
Common Device Name
VENACURE 400 MICRON PERFORATOR AND ACCESSORY VEIN ABLATION KIT
Manufacturer (Section D)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key18825296
Report Number1319211-2024-00032
Device Sequence Number7749236
Product Code GEX
UDI-Device IdentifierH787EVLTPVAKUS5
UDI-PublicH787EVLTPVAKUS5
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K041957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberH787EVLTPVAKUS5
Device Lot Number5806941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2024
Supplement Date Received by Manufacturer04/18/2024
Initial Report FDA Received Date03/04/2024
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/11/2023
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-