MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTERLOCK; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number 83779
Medical Device Problem Codes Stretched (1601); Failure to Advance (2524); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.Visual inspection was performed and it was observed that the main coil and the pusher wire were returned.It was observed that the main coil was bent and stretched at the coil arm and primary coil section.Also, the zap tip was detached.No more damages were observed during visual inspection.The functional inspection could not be performed due to the main coil and the pusher wire are not interlocking.Dimensional inspection of the main coil revealed the components were within specifications.No other issues were identified during the product analysis.
 
Event or Problem Description
Reportable based on device analysis completed on (b)(6) 2024.It was reported that coil advancing and stretching issues occurred.The stenosed target lesion was located in the internal iliac artery.A 4mm x 15cm interlock coil was selected for embolization of the internal iliac aneurysm.During the procedure, the coil was delivered by a bsc microcatheter for embolization, and heparin saline perfusion was performed.However, clear resistance was felt, and the coil could not be advanced when it reached the 2-3cm segment of the catheter.Consequently, it was found that the coil was stretched upon removal from the catheter.The procedure was completed with another of the same device.No complications were reported, and the patient was stable post-procedure.However, device analysis revealed a detached zap tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18825704
Report Number2124215-2024-08846
Device Sequence Number3076029
Product Code KRD
UDI-Device Identifier08714729764960
UDI-Public08714729764960
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132578
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2026
Device Model Number83779
Device Catalogue Number83779
Device Lot Number0031454080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2024
Initial Report FDA Received Date03/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexFemale
-
-