• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/S NPOR FEM RT SZ 4; KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC*SIGMA C/S NPOR FEM RT SZ 4; KNEE FEMORAL Back to Search Results
Catalog Number 960054
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Joint Laxity (4526); Unspecified Tissue Injury (4559)
Date of Event 01/25/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Hcp sent a medical info request to team: "i have a patient in the clinic who has developed symptoms since several weeks after implantation of a pfc modular plus prosthesis.Surgical date (b)(6) 2008: pfc modular right knee: femur cemented, size 4, tibia cemented, size 3, plateau height 10 mm stabilized; patellum shaping.The patient now presents and the images show the inlay as worn/defective with a defect on the dorsal medial tibial console and ventral subluxation of the tibia to the femur.It is clear that there will be a certain amount of metallosis with a dorsally rubbed tibial component.As the manufacturer, is it still permissible to change the inlay alone in the case of a dorsally impaired tibial component?".
 
Additional Manufacturer Narrative
Product complaint (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information was received.No allegation against patella.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, "hcp sent a medical info request to team: "i have a patient in the clinic who has developed symptoms since several weeks after implantation of a pfc modular plus prosthesis.Surgical date (b)(6) 2008: pfc modular right knee: femur cemented, size 4, tibia cemented, size 3, plateau height 10 mm stabilized; patellum shaping.The patient now presents and the images show the inlay as worn/defective with a defect on the dorsal medial tibial console and ventral subluxation of the tibia to the femur.It is clear that there will be a certain amount of metallosis with a dorsally rubbed tibial component.As the manufacturer, is it still permissible to change the inlay alone in the case of a dorsally impaired tibial component?"".The product was not returned to depuy synthes, however photos were provided for review.See attachment ((b)(4) x-rays).The x-ray investigation revealed nothing indicative of a device nonconformance.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was unconfirmed as the observed condition of the pfc*sigma c/s npor fem rt sz 4 would not contribute to the reported adverse event.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: 1) quantity manufactured: (b)(4).2) date of manufacture: 23-oct-2007.3) any anomalies or deviations identified in dhr: no.4) expiry date: 5/2/2014.5) ifu reference: ifu-0902-00-252.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18825866
Report Number1818910-2024-04912
Device Sequence Number18089144
Product Code JWH
UDI-Device Identifier10603295232384
UDI-Public10603295232384
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K971189
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/02/2014
Device Catalogue Number960054
Device Lot Number2479227
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2024
Supplement Date Received by Manufacturer03/13/2024
05/17/2024
Initial Report FDA Received Date03/04/2024
Supplement Report FDA Received Date03/19/2024
05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/23/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PFC SIGMA STAB INS SZ3 10MM; PFC SIGMA TIB TRAY CEM SZ3; PFC*SIGMA/OV DOME PAT 3 PEG,32; UNKNOWN KNEE TIBIAL INSERT; UNKNOWN KNEE TIBIAL TRAY
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-