| Model Number |
1156.0010 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being resubmitted for an incident that occurred earlier.Evaluation of the returned locking caps and rod was performed.The threads had signs of wear to due possible edge loading.It's possible that micro-motion within the construct due to the construct pulling out of bone, as well as extreme bends seen in the rod causing the locking caps to be edge loaded and not locking appropriately may have contributedto the reported issue.However, the exact cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the creo dlx locking cap had loosened approximately one week post-operatively.
|
| |
|