Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the administration of the second dose of a split dose case of sir-spheres treatment, dr.(b)(6) was no longer able to inject through either the b or d lines.Dr.(b)(6), who was proctoring dr.(b)(6) , was also present and assisted in trouble shooting to attempt to clear the line.The line was pulled and case was completed.The entire system was sent to be decayed.Post survey showed approximately 45% of the dose was delivered to patient prior to clogging (prescribed dose.3gbq, administered was.14 gbq).Non-sirtex equipment used was a sim 2 catheter into the celiac trunk and 2.4fr.Progreat into the distal portion of segment 7.Nuc med is holding the system until it's decayed to background and will ship back for quality to check the system to see if the issue was the siros tubing or a clogged/kinked microcatheter.Also of note, inside of siros box was hot treatment.The d port connection going into the d-vial was cracked.Suspect due to pressure build up within the system when dr.Hazelton was no longer able to inject.
|
| |
|
Additional Manufacturer Narrative
|
|
The described issue stated in the incident report could not be reproduced during the investigation.Water flowed freely through the lines of the device in all configurations, and no pressure was felt.The device history records (dhr) review showed no abnormalities during the manufacturing process and that the product met the production acceptance criteria per established manufacturing production instructions.The reported incident was likely attributed to operator technique, or a connection during the operation that could not be replicated during the investigation of the device.The root cause of the incident cannot be confirmed.
|
| |
|
Event or Problem Description
|
|
During the administration of the second dose of a split dose case of sir-spheres treatment, dr.(b)(6) was no longer able to inject through either the b or d lines.Dr.(b)(6), who was proctoring dr.(b)(6), was also present and assisted in trouble shooting to attempt to clear the line.The line was pulled and case was completed.The entire system was sent to be decayed.Post survey showed approximately 45% of the dose was delivered to patient prior to clogging (prescribed dose.3gbq, administered was.14 gbq).Non-sirtex equipment used was a sim 2 catheter into the celiac trunk and 2.4fr.Progreat into the distal portion of segment 7.Nuc med is holding the system until it's decayed to background and will ship back for quality to check the system to see if the issue was the siros tubing or a clogged/kinked microcatheter.Also of note, inside of siros box was hot treatment.The d port connection going into the d-vial was cracked.Suspect due to pressure build up within the system when dr.(b)(6) was no longer able to inject.
|
| |
|
Search Alerts/Recalls
|