|
It was reported that, during an internal fixation procedure, the tip of one (1) 7.0 compression screw drill broke off intraoperatively and could not be retrieved.The piece was left in the patient.The procedure was resumed, without any delay, using the same device.Patients' current health status is unknown.
|
|
H3, h6: the associated device was returned and evaluated.The visual inspection revealed that the tip of the device broke off.The broken piece was not returned.The device shows signs of significant wear and use.This inspection was conducted by naked eye.The clinical/medical investigation concluded that, as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the broken drill tip.The procedure was resumed, without any delay, using the same device.Although unlikely, the potential for corrosion and local irritation/migration of the drill tip could not be ruled out.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of instructions for care, maintenance, cleaning and sterilization of smith & nephew orthopedics devices for single use devices revealed that as the devices are sold both nonsterile and sterile and often removed from their original packaging to be placed in a containment device they should be cleaned and/or inspected prior to sterilization.The device should not be reprocessed or reused if comes in contact with blood, tissue or bodily fluids.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This is a single use device, surgical technique or damage from misuse are likely potential factors that could contribute to the reported event.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|