| Model Number |
MAJ-2127 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, the connectors on the leak test air tube kept breaking.The issue occurred during reprocessing.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.However, it is likely that the user applied external stress in the incorrect direction to the lock lever of the connecting tube, which broke the lever and the tube was unable to be connected.The user can detect the event by conducting inspection as follows per the instructions for use (ifu): "9.Preparation and inspection: 1.Visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.2.Move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken." olympus will continue to monitor field performance for this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation.The device history record was unable to be reviewed since the exact manufacture date is unknown.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.The device was returned to olympus for inspection, and the customer's reportable malfunction was confirmed.Olympus will continue to monitor field performance for this device.
|
| |
|