Catalog Number SAGL2022
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Subluxation (4525)
Date of Event 02/12/2024
Type of Reportable Event Serious Injury
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following section was corrected: h4.Reported event was confirmed by review of medical records.Review of the available records identified the following: on (b)(6) 2024 pt reported that they "slept wrong".Initial appt noted pain, reduced rom, posterior subluxation, and negative labs.A displaced coracoid tip fracture was also noted.Implants aligned and well fixed.Manipulation (closed reduction) was attempted with no success, sent to referral/revising surgeon.On (b)(6) 2024 consult with revising surgeon noted no history of trauma, limited rom, axillary nerve slight paresthesia, drop arm positive, with joint dislocation.On (b)(6) 2024, pt revised to reverse tsa.Revision notes posterior dislocation, muscular tissues dense and scarred.Humeral head centricity rotated posterior inferior, stem appeared to be degrees of retroversion.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Device is used for treatment.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following section was corrected: e1.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device has been discarded by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient's left glenoid was revised two months post-initial procedure due to pain, limited rom, bone fracture, and subluxation.No additional patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3 PEG MONO GLEN SZ 2
Common Device Name
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18827529
Report Number0001822565-2024-00694
Device Sequence Number3008681
Product Code KWS
UDI-Device Identifier00889024644670
UDI-Public(01)00889024644670(17)280627(10)66074643
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K191814
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSAGL2022
Device Lot Number66074643
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/22/2024
Supplement Date Received by Manufacturer02/29/2024
06/24/2024
Initial Report FDA Received Date03/04/2024
Supplement Report FDA Received Date03/18/2024
07/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight65 KG
-
-