MEDOS INTERNATIONAL SARL ENTERPRISE2 4MMX30MM NO TIP; INTRACRANIAL NEUROVASCULAR STENT Back to Search Results
Catalog Number ENF403000
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Information regarding patient identifier, date of birth, age, gender, weight, race, and ethnicity were not provided.Section e.1: the name, phone and email address of the initial reporter are not available / reported.Section h.3: the device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.The product analysis team reviewed the photo included in the complaint.The review is documented below.[photo review]: the photo included in the complaint shows the detached stent inside the microcatheter hub.No appearance of damage was observed.The delivery wire was noted without damage.The introducer is not observed in the photo and no further damages could be noted.Residues of dried blood were noted in the hub of the microcatheter.The issue regarding a stent being pre-deployed in the hub of the concomitant microcatheter was confirmed based on the detached condition of the stent.This investigation was performed based only on the photo provided.If the product is received after this investigation, an assessment will be performed as per the conditions of the device returned.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 7057096.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
The healthcare professional reported that during a vascular stent placement procedure, the 4.0mm dia x 30mm with no tip enterprise®2 (enf403000 / 7057096) was pre-deployed in the hub of the concomitant 150cm x 5cm prowler select plus microcatheter (606s255x / 30871574).There was no report of any negative patient impact.A photo was included in the complaint.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.(b)(4).The purpose of this mdr submission is to report that multiple attempts to obtain additional information related to the procedure, the reported device issue, and to obtain the product for analysis were unsuccessful.If additional information is available at a later date and the product is returned, the file will be updated accordingly.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 7057096.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.With the information available and without the complaint product available to be returned for analysis, the reported issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Updated sections: b.4, g.3, g.6.H.2, h.3, h.6, and h.10.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report that the product analysis lab received the complaint device on (b)(6) 2024.A supplemental 3500a report will be submitted once the product investigation has been completed.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to report the investigational finding of the returned device.The complaint product was returned and received for evaluation and analysis.The investigation is documented below.Investigation summary: a non-sterile 4.0mm dia x 30mm with no tip enterprise®2 was received contained in the decontamination pouch.Visual inspection was performed.It was observed that the stent was returned detached inside the hub of the microcatheter.The delivery wire and the introducer were found in good conditions.The stent component was removed and inspected under the microscope.Under magnification, it was observed to be in good condition.There was no structural damage (i.E., no broken struts, no kinks).Both ends of the stent were observed completely flared.The delivery wire underwent dimensional analysis.All measurements were within specifications, including those specifications that control the attachment and delivery of the stent.Therefore, device failure is not suspected to be a contributing factor.The reported issue documented in the complaint regarding a stent being released in the microcatheter hub was confirmed since the stent was noted as already separated from the delivery system.However, the dimensional analysis performed showed no issues that could have resulted in the premature detachment of the stent.With the information available, the root cause of the failure encountered cannot be determined.It is possible that clinical and procedural factors, including device manipulation, may have contributed to the reported failure.There is no indication that the issues reported in the complaint result from a defect inherently related to the enterprise device.Lake region medical performed a review of the device history records relative to the manufacturing, inspection, and packaging of the lot 7057096.The history record indicates this product was final inspection tested at lake region medical and was determined to be acceptable.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) does contain the following recommendations: do not partially deploy the stent from the introducer.Confirm that the delivery wire does not move relative to the introducer during the removal of the cerenovus enterprise vascular reconstruction device and delivery system from the dispenser hoop.Confirm the tip of the delivery wire is entirely within the introducer.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERPRISE2 4MMX30MM NO TIP
Common Device Name
INTRACRANIAL NEUROVASCULAR STENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CERENOVUS INC
325 paramount dr
raynham MA 02767
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key18828314
Report Number3008114965-2024-00248
Device Sequence Number12709412
Product Code NJE
UDI-Device Identifier10886704075387
UDI-Public10886704075387
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
H60001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Catalogue NumberENF403000
Device Lot Number7057096
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2024
Supplement Date Received by Manufacturer04/04/2024
04/22/2024
05/06/2024
Initial Report FDA Received Date03/04/2024
Supplement Report FDA Received Date04/04/2024
04/22/2024
05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROWLER SELECT PLUS 150/5CM.
Patient SexUnknown
-
-