| Model Number |
KINGFIR SUTR RETRVR /TISS GRSPR W/WB |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 02/16/2024, it was reported by a distributor via (b)(4) that an ar-13970w suture retrievers clamp broke.This occurred during a case.No additional information was provided, and additional information has been asked.
|
| |
|
Event or Problem Description
|
|
Additional information was provided on (b)(6) 2024, stating that this event occurred during a meniscus arthroscopy procedure on (b)(6) 2024.The device broke while in use, and all fragments were retrieved.Another clamp was used to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.One unpackaged ar-13970w suture retriever lot number: 11818633 was received for evaluation.Visual inspection noted that the jaw of the device would not completely close.Functional testing was performed with a test sheet and it was observed that the ar-13970w suture retriever would not grasp.The most likely cause can be attributed to misuse/mishandling due to the damage to the device.Refer to investigation photos.
|
| |
|