| Model Number |
561-0510 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Arthralgia (2355)
|
| Date of Event |
02/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
On (b)(6) 2024, a femoral sleeve construct was revised.After exposing the knee, it was noted that the junction box slid anterior and posterior on the femoral component.The morse taper of the junction box into the sleeve was intact and strong.
|
| |
|
Additional Manufacturer Narrative
|
|
A dhr review identified no issues that could have caused or contributed to the reported event.It is believed that the junction box loosened due to fatigue loading in the knee.The available evidence does not lead to a clear conclusion about the cause of the reported event.
|
| |
|