• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 37603
Medical Device Problem Codes Charging Problem (2892); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient had a damaged desktop charger (dtc) cord.The dtc was damaged "about a month ago." then, "about saturday," they noticed the recharger did not seem to take a charge from the dtc unless they jiggled the dtc.On that same day, the caller also noticed that the dtc connector pin seemed loose.During the call, the caller noted that the dtc connector pin did not seem to be as loose.
 
Additional Manufacturer Narrative
Continuation of d10: product id 37761 lot# serial# (b)(6), product type recharger section d information references the main component of the system.Other relevant device(s) are: product id: 37761, serial/lot #: (b)(6).H3: analysis of the 37761 desktop charger (dtc) (s/n (b)(6) revealed broken connector pins, and worn rubber feet.This regulatory report is being submitted as part of a retrospective review as part of remediation plan 411.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18831403
Report Number3004209178-2024-06229
Device Sequence Number19594996
Product Code MHY
UDI-Device Identifier00613994761071
UDI-Public00613994761071
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2016
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2022
Initial Report FDA Received Date03/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H11...."
Patient Age81 YR
Patient SexMale
-
-