It was reported that during an unspecified number of colon polypectomy procedures, the plastic parts of the coil sheath and stopper of the subject device were bent when the unidentified scope was angled, and the handle on the hand side was bent as well while strangling a polyp and pushing out the handle to release the snare.The metal part of the snare broke and could not be released, and the device had to be cut with nippers and other unspecified tools.The procedures were completed with the same device.The customer could not determine the exact dates and number of times the issue occurred.There were no reports of patient harm.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to the initial with information inadvertently left out (g2).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and past investigation results, a likely mechanism causing the reported events could be one of the following: mechanism 1: 1) the loop was surrounding the body tissue, and it was temporarily ligated by pulling the slider.2) the tube sheath was pushed out, and the distal end of the coil sheath went into the tube sheath.3) an attempt was made to detach the loop in state of above description 2).Therefore, the loop detached from the hook in the tube.4) while the hook was extending from the coil sheath, the loop moved towards the proximal side and went into the coil sheath.5) the hook was pulled.This caused the hook and the loop to retract into the coil sheath together.As a result, the loop and the hook got stuck inside the coil and could not move.6) since the loop and the hook got stuck together inside the coil, the loop did not detach when the slider was pulled.7) the slider was forcefully operated in state of ¿6¿ description.This had caused the operating pipe to bend and to break.Mechanism 2: 1) the sheath was bent near the handle.2) the operating wire could not move because sliding resistance between the sheath and the operating wire increased.3) the operating pipe deformed and broke because the slider was forcefully operated.4) due to above, the loop could not detach from the hook.However, the subject device was not returned, and the root cause of the suggested event could not be determined.The event can be prevented by following the instructions for use which state: ¿ do not strike or crush the coil sheath during operation.Doing so can damage the distal end of the coil sheath, which could make it impossible to detach the loop after ligation.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.¿ do not remove the loop from the hook while the coil sheath is not extended from the tube sheath.Otherwise, the loop may be tangled with the hook and become impossible to be removed.In this case, refer to section12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.¿ do not hold the loop with the distal end of the tube sheath while the loop is surrounding the tissue.Otherwise, when the tissue is ligated, the loop may be detached from the hook in the tube sheath and tangled with the hook.That may make the loop impossible to be removed.In this case, refer to section 12, ¿emergency treatment¿ and as shown ¿equipment to be used in an emergency¿ on page 3 in this manual.¿ never use excessive force to operate the instrument.This could damage the instrument.Olympus will continue to monitor field performance for this device.
|