• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE PRESERVE STEM 6MM; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE PRESERVE STEM 6MM; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 300-30-06
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Periprosthetic fracture (4902)
Date of Event 01/24/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(d10) concomitant device(s): 320-38-00 - 145-deg pe 38mm hum liner +0: a550004 320-10-00 - equinoxe reverse tray adapter plate tray +0: a285236 320-06-38 - glenosphere 38mm: a860948 320-15-07 - sup/post aug plate, l rs glenoid baseplate: a826082 315-35-00 - glnd kwire: a585081 315-35-00 - glnd kwire: a820684 320-15-05 - eq rev locking screw: a603142 320-20-00 - eq reverse torque defining screw kit: a855998 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm: s509134 320-20-34 - eq rev compress screw lck cap kit, 4.5 x 34mm: s502520 320-20-34 - eq rev compress screw lck cap kit, 4.5 x 34mm: s502614 521-78-31 - threaded pin size 2.6 collarless 2pk: s503241 531-55-88 - ergo gps 3.2mm drill kit sterile: a762816 (h3) pending evaluation.
 
Event or Problem Description
As reported by the equinoxe shoulder study, the 76-year-old female patient had a left tsa on (b)(6) 2024 and presented with humeral fracture - midshaft, intra-operatively on (b)(6) 2024.Intra-op humeral calcar fracture during reduction.The outcome of this event is considered continuing, with the action taken of other - cerclage cable around humerus.The case report form indicates that this event is unlikely related to the device and definitely related to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c, d, e, f, g.The reason for the intraoperative bone fracture reported cannot be conclusively determined but may be related to a patient condition or high forces during broaching, reaming, exposure, or retraction.However, this cannot be confirmed based on the information provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE PRESERVE STEM 6MM
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
miguel sosa
2320 nw 66th ct.
gainesville, FL 32653
3523771140
MDR Report Key18833091
Report Number1038671-2024-00387
Device Sequence Number7485717
Product Code PHX
UDI-Device Identifier10885862515742
UDI-Public10885862515742
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K162726
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number300-30-06
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer08/04/2025
Initial Report FDA Received Date03/04/2024
Supplement Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/07/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient Age76 YR
Patient SexFemale
Patient Weight104 KG
Patient EthnicityNon Hispanic
Patient RaceBlack Or African American
-
-