|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additionally, and correction to the initial with information inadvertently left out (d4-lot number and d9) and to provide an update to fields (h3).The subject device was manufactured on april 2023 based on the provided 3 digit lot information "34k".However, the exact date is unknown.The device was evaluated by olympus, and the tube was broken at 15 mm from the distal end of the insertion portion.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely event 1 " the patient ended up going to cardio-thoracic in the rie for treatment" due to the device malfunction.Additionally, it is possible the balloon fell out because the tubing ruptured.The following mechanisms may have caused the tube to rupture.1) the tip of the tube became fixed during foreign matter collection.2) the tube was pulled further while it was fixed.3) a load was applied to the tube, causing it to break.The following mechanisms could be responsible for balloon dislodgement from the fixation site.1) a force beyond the strength resistance was applied to the balloon when retrieving foreign materials.2) the fixed area at the rear end of the balloon broke, and it was displaced to the distal end.However, the root cause of the reported event could not be determined.The event can be prevented by following the instructions for use which state: ·do not use the balloon catheter to retrieve a calculus that is larger than the fistula or lumen size.Doing so could cause patient injury such as bleeding or mucous membrane damage.It could also burst the balloon or cause the balloon to become stuck in the fistula or lumen, resulting in the distal end of the instrument breaking off and remaining inside the fistula or lumen.·the volume of the air in the balloon should not exceed the parameters specified in the tables in chapter 8, ¿specifications¿.Otherwise, the balloon may burst or may not deflate properly.When the balloon does not deflate, do not operate the balloon catheter with excessive force.Otherwise, the distal end of the instrument may break off and could remain inside the patient.·do not insert the instrument into the endoscope unless you have a clear endoscopic field of view.If you cannot see the distal end of the insertion portion in the endoscopic field of view or in the x-ray images, do not use it.This could cause patient injury, such as punctures, bleeding, or mucous membrane damage.It may also damage the endoscope and/or instrument.·do not operate the instrument abruptly or with excessive force when retrieving a foreign object.Otherwise, patient injury such as bleeding, mucous membrane damage, or edema may occur.·if the balloon cannot be deflated, withdraw it using an appropriate method determined by the specialist.If an excessive force is applied on the balloon catheter when the balloon cannot be deflated, patient injury such as bleeding, mucous membrane damage, or edema may occur.It could also cause the balloon to burst or the distal end of the instrument to break off, which could remain inside the patient.·do not withdraw the instrument from the endoscope if the forceps elevator is up.This could damage the instrument.Olympus will continue to monitor field performance for this device.
|