| Model Number |
2315-3213, 2115-1330 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient reported that they got out of a chair and experienced pain.They indicated no fall or trauma.Previous ustar ii patient presented with pain and instability.X-ray imaging revealed dislocation of the femoral component and tibial insert yoke and a revision was scheduled.During surgery, surgeon noted the insert screw connecting the femoral component insert and yoke had bent.Original surgery was on (b)(6) 2023 and revision was conducted on (b)(6) 2024.Revision surgery was successfully completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Receiving the awareness of this case from importer, investigation into the cause of the adverse event is in progress.
|
| |
|
Additional Manufacturer Narrative
|
|
According to the investigation, it is confirmed from the product manufacturing inspection records that the dimensions of the abnormal device met the manufacturing tolerance.From the observation of the returned device, there is no evidence of assembly between the screw taper and the internal taper of the adapter, moreover, significant pressure marks are present on the end face of taper of the screw.Based on the comprehensive analysis, it is inferred that the root cause of this case is misalignment during the assembly of the adapter and screw, ultimately leading to implant dislocation.
|
| |
|
Event or Problem Description
|
|
The patient reported that they got out of a chair and experienced pain.They indicated no fall or trauma.Previous ustar ii patient presented with pain and instability.X-ray imaging revealed dislocation of the femoral component and tibial insert yoke and a revision was scheduled.During surgery, surgeon noted the insert screw connecting the femoral component insert and yoke had bent.Original surgery was on (b)(6) 2023 and revision was conducted on (b)(6) , 2024.Revision surgery was successfully completed.
|
| |
|