Model Number 1119.0110
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being resubmitted for an incident that occurred earlier.The device was returned for evaluation.Initial observation shows minor signs of wear to the thread features and bottom contacting surface of the locking caps.This is consistent with inadequate final tightening of the implants.The imaging provided confirmed several caps back-out at the top of the construct, where the rods and locking caps had dissociated from the screws.It is possible that the locking caps were not sufficiently final tightened due to the lack of surface and thread wear, however, the exact cause of the reported issue could not be determined.
 
Event or Problem Description
It was reported there was a revision due to creo amp threaded polyaxial screw heads backing out of a construct post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREO STABILIZATION SYSTEM
Common Device Name
CREO AMP THREADED POLYAXIAL TULIP
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key18834300
Report Number3004142400-2023-00084
Device Sequence Number18001990
Product Code MNH
UDI-Device Identifier00889095059656
UDI-Public00889095059656
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K124058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1119.0110
Device Lot NumberBAA681LG/BAA276HG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/04/2023
Initial Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-