ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL; INTRAOCULAR LENS Back to Search Results
Model Number DFT415
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A other healthcare professional reported that during the intraocular lens (iol) implant procedure, when lens was inserted into eye it appeared to be rigid and when implanted noticed stretch mark on lens.The surgeon removed and replaced the lens with new one.Additional information has been received that there was no patient harm reported.
 
Additional Manufacturer Narrative
The lens was returned for evaluation.Solution and blood are dried on the lens.One haptic is broken in the gusset area and returned loose in the lens case.The opposite haptic is broken in the distal area (not returned).The optic has been cut into pieces, typical of removal from the eye.Only two pieces of the optic have been returned.The lens was cleaned for further evaluation.Scratches and cracks are observed on the portions of the lens that were returned.Product history records were reviewed and the documentation indicated the product met release criteria.The file indicates the use of a qualified cartridge and handpiece.The file also indicates the use of a non-qualified viscoelastic.Lens damage was observed.The root cause for the reported complaint may be related to a failure to follow the instructions (ifu).The viscoelastic indicated is not qualified for the lens/monarch combination used.Company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18835101
Report Number1119421-2024-00405
Device Sequence Number7132851
Product Code POE
UDI-Device Identifier00380652421881
UDI-Public00380652421881
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2026
Device Model NumberDFT415
Device Lot Number15188929
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2024
Supplement Date Received by Manufacturer03/21/2024
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HYALURONATE; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Patient SexFemale
-
-