| Model Number |
87500 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
01/19/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A patient allegedly experienced postoperative 2nd degree burns to the abdomen and right leg with the use of the 3m¿ bair hugger¿ patient adjustable warming unit 87500.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated, and no fault was found with the device related to this event.The device passed the critical safety over-temperature (ot) test and shuts off when the output air temperature exceeds 47°c.No damage or degradation observed to pcba, heater, or blower critical components.3m will continue to monitor.
|
| |
|