Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D2b: additional device product codes: hwc d9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that during surgery on (b)(6) 2024, the surgeon was performing a dual plating of the distal humerus, and multiple variable angle locking screws would not lock into each plate.Standard operating procedure was performed for each screw that consisted of using a cone guide for drilling followed by measuring and insertion of the screw with torque limiter.There were no patient outcomes reported.The surgery was completed successfully with a delay of 15 minutes.This report involves one 2.7mm/3.5mm va-lcp ext medl dstl hum pl 1h/rt/72mm-short.This is report 2 of 4 for (b)(4).
|