MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA TM CRUCIATE RETAINING NARROW POROUS FEMORAL COMPONENT RIGHT SIZE 10; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA TM CRUCIATE RETAINING NARROW POROUS FEMORAL COMPONENT RIGHT SIZE 10; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during knee arthroplasty that when the femoral implant was opened and the first layer of film peeled back, a hair was identified on the rim of the packaging for the implant in the sterile area.The surgeon was notified and a new femoral implant was used to complete the procedure.No adverse events were reported as a result of this malfunction.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.Visual review of the provided photo identified a hair-like debris on the blister flange.Visual evaluation of the returned product could not locate the reported hair-like debris.No damage to the packaging or device was noted.As the product was returned opened, the source of the debris cannot be confirmed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause or condition of the device when it left zimmer biomet cannot be determined, as the product was returned opened.This complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA TM CRUCIATE RETAINING NARROW POROUS FEMORAL COMPONENT RIGHT SIZE 10
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18836707
Report Number0001822565-2024-00749
Device Sequence Number7181467
Product Code MBH
UDI-Device Identifier00889024470149
UDI-Public(01)00889024470149(17)291130(10)64553656
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42502206802
Device Lot Number64553656
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2024
Supplement Date Received by Manufacturer06/06/2024
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date06/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/22/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-