This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the observed foreign material nozzle and on the objective lens adhesive could not be identified and a definitive root cause of the issue could not be determined, as there was no deformation observed that might result in the retention of foreign material and no additional facility device reprocessing information was reported or available, however, the issue was likely the result of insufficient device cleaning/reprocessing.The issue may be detected/prevented by following the instructions for use sections below: gastrointestinal videoscope,gif-lv1,colonovideoscope,cf-lv1l,cf-lv1i,operation manual chapter 3 preparation and inspection.Gastrointestinal videoscope,gif-lv1,colonovideoscope,cf-lv1l,cf-lv1i,reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).Olympus will continue to monitor field performance for this device.
|