MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, EXTRA LORDOTIC, 3-LEVEL, 54MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, EXTRA LORDOTIC, 3-LEVEL, 54MM Back to Search Results
Model Number 1194.3154
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being resubmitted for an incident that occurred earlier.The device was returned for evaluation.Initial observation shows one broken slider and one broken screw.The imaging providd showed one back out screw and one broken screw, both at the bottom level.No determinations could be made as to the cause of the reported issue.The following sections have been updated: b4, e1, h2, h6, h10.
 
Event or Problem Description
It was reported that a revision surgery was needed to replace a screw that backed out of a resonate plate post-operatively with subsequent screw breakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE
Common Device Name
RESONATE ANTERIOR CERVICAL PLATE, EXTRA LORDOTIC, 3-LEVEL, 54MM
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key18837525
Report Number3004142400-2023-00109
Device Sequence Number20716030
Product Code KWQ
UDI-Device Identifier00193982155551
UDI-Public00193982155551
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K192314
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1194.3154
Device Lot NumberETX330AA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2023
Initial Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-