| Model Number |
1194.2028 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being resubmitted for an incident that occurred earlier.The device was not available for evaluation as it remains in the patient.The immediate post-operative imaging showed all the screws fully seated.The post-operative imaging taken after six months the surgery show the bone screw head slightly protruding from the plate and a small piece of foreign material next to it.It is unclear if the small piece is a part of the plate, a slider or something else.No determinations could be made as to the cause of the reported issue.
|
| |
|
Event or Problem Description
|
|
It was reported that screws are backing out of a resonate plate post operatively.
|
| |
|