• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Catalog Number BP-05501
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/13/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that "while attempting to remove epidural catheter, the catheter broke.It was determined that surgery is needed to remove the flextip plus catheter due to a concern the catheter might migrate if left in patient.Surgery scheduled for week of (b)(6) 2024.Patient as i know it is fine, but still need surgery to remove the catheter and that has not happened yet".At the time of this report, the customer has not returned our requests for additional information.
 
Additional Manufacturer Narrative
Qn# (b)(4).
 
Additional Manufacturer Narrative
Qn# (b)(4).Additional information received indicates "patient required laminectomy to remove retained catheter".It was also reported that the patient did not experience any signs or symptoms due to this event, and the condition of the patient was "stable".The actual device was not returned; however, the customer did provide a photo that appears to show a separated catheter.A device history record review was performed on the epidural catheter with no relevant findings.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter with no relevant findings.The customer did provide a photo that appears to show a separated catheter; however, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.Other remarks: n/a, corrected data: n/a.
 
Event or Problem Description
It was reported that "while attempting to remove epidural catheter, the catheter broke.It was determined that surgery is needed to remove the flextip plus catheter due to a concern the catheter might migrate if left in patient.Surgery scheduled for week of (b)(6) 2024.Patient as i know it is fine, but still need surgery to remove the catheter and that has not happened yet".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18837690
Report Number9680794-2024-00217
Device Sequence Number15238974
Product Code BSO
UDI-Device Identifier50801902007457
UDI-Public50801902007457
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberBP-05501
Device Lot Number33F23I0385
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/14/2024
Supplement Date Received by Manufacturer04/01/2024
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date04/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexFemale
Patient Weight64 KG
-
-