PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 20E DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 20E
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted stryker to report that their device would not complete its initial boot-up cycle. in this state the device may not be able to deliver defibrillation therapy, if needed.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
A stryker service representative performed an evaluation of the customer¿s device and was able to verify the reported issue.The technician replaced the sc and ui pcba's to repair device.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.Stryker further evaluated the replaced part at the product assessment center (pac) and was not able to duplicate the reported issue.The cause of the reported issue was faulty sc and ui pcba's, but a further root cause could not be determined.
 
Event or Problem Description
The customer contacted stryker to report that their device would not complete its initial boot-up cycle. in this state the device may not be able to deliver defibrillation therapy, if needed.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20E DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key18837780
Report Number0003015876-2024-00488
Device Sequence Number20716033
Product Code MKJ
UDI-Device Identifier00883873853863
UDI-Public00883873853863
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P160026
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number20E
Device Catalogue Number99507-000122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by Manufacturer07/31/2024
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date08/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-