| Model Number |
20E |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not complete its initial boot-up cycle. in this state the device may not be able to deliver defibrillation therapy, if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
A stryker service representative performed an evaluation of the customer¿s device and was able to verify the reported issue.The technician replaced the sc and ui pcba's to repair device.Proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.Stryker further evaluated the replaced part at the product assessment center (pac) and was not able to duplicate the reported issue.The cause of the reported issue was faulty sc and ui pcba's, but a further root cause could not be determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not complete its initial boot-up cycle. in this state the device may not be able to deliver defibrillation therapy, if needed.There was no patient use associated with the reported event.
|
| |
|