• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, 2-LEVEL, 32MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. RESONATE; RESONATE ANTERIOR CERVICAL PLATE, 2-LEVEL, 32MM Back to Search Results
Model Number 1194.2032
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a screw is backing out of the resonate plate post-operatively.
 
Additional Manufacturer Narrative
This report is being resubmitted for an incident that occurred earlier.The device was not available for evaluation as it remains in the patient.The immediate post-operative image all screws appear to be fully seated.The follow-up images show the head of a bottom screw is partially protruding from the plate.It is unclear whether the cause was a damaged slider or if the screw was not fully seated in the plate.No determinations could be made as to the cause of the reported issue.The following sections are been updated: b4, e1, h2, h6, h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE
Common Device Name
RESONATE ANTERIOR CERVICAL PLATE, 2-LEVEL, 32MM
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key18838074
Report Number3004142400-2023-00138
Device Sequence Number20704528
Product Code KWQ
UDI-Device Identifier00193982154837
UDI-Public00193982154837
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K192314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1194.2032
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/09/2023
Initial Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
Patient SexFemale
Patient RaceBlack Or African American
-
-