• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB 25G BI-B POSTERIOR ELITE W WF; UNIT, PHACOEMULSIFICATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB 25G BI-B POSTERIOR ELITE W WF; UNIT, PHACOEMULSIFICATION Back to Search Results
Model Number SE5425WVB
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One 25ga bi-blade vitrectomy cutter was returned in a plastic bag, along with an se5425wvb pack label from lot x5120.The tip protector and the rest of the pack components were not returned.The needle was bent.The port window was in the opened position with debris and solutions visible in and around the port and on the outer needle shaft.There was fluid in the tubing.This cutter looked used.The back cap was correctly aligned with the vent hole in the side of the cutter body.The connectors were inspected and no damage was found.A functional test was performed using a stellaris elite system.The cutter did not cut, but did aspirate.It should be noted that a bent needle could contribute to poor cutting performance.Due to evidence of use, and the returned condition without the tip protector, the cause of the bent needle cannot be determined.The lot history, trend analysis, risk analysis and or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action required.
 
Event or Problem Description
A user facility reported that the vitrectomy cutter stopped working.There was no medical intervention required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
25G BI-B POSTERIOR ELITE W WF
Common Device Name
UNIT, PHACOEMULSIFICATION
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
3365 tree court industrial blv
st. louis MO 63122
Manufacturer Contact
juli moore
6362263220
MDR Report Key18838627
Report Number0001313525-2024-70061
Device Sequence Number17925401
Product Code HQC
UDI-Device Identifier00757770059775
UDI-Public(01)00757770059775(17)250317
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K170052
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/17/2025
Device Model NumberSE5425WVB
Device Lot NumberX5120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/05/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
-
-