| Catalog Number |
826653 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
A facility reported a microsensor (b)(6) was implanted on (b)(6) 2024 at 6:00pm, and around 10:30 pm, the microsensor was found leaking a moderate amount of cerebrospinal fluid (csf), the surgical dressing was wet.On january 10, 2024, the physician found that there was no more cerebrospinal fluid (csf) leaking, and retrieved the microsensor.The sensor was not replaced and the patient did not experience any signs and symptoms due to csf leaking.The patient is in good condition.The monitor used was icp express monitor (b)(6), cable used was icp express cable (b)(6).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The microsensor (id 826653) was not returned for evaluation as the product was discarded as per customer; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause(s) of the reported issue could not be determined however, a potential cause of the reported failure may be related to an unidentified unsecure connection during placement.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
| |
|
Search Alerts/Recalls
|