| Model Number |
WA22503D |
| Medical Device Problem Codes |
Material Rupture (1546); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using the high-frequency resection electrode during a hysteroscopy resection (patient under spinal anesthesia) therapeutic procedure, it was observed that there was a rupture to the terminal part of the single-use resection loop.The patient buttocks abruptly moved and she experienced pain without an error message being seen on the generator.The patient then complained of a metallic taste in her mouth.The procedure was stopped, and when the loop was removed, there was a rupture in the terminal part of the loop.There was no medical intervention required as a result of the reported device problem.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the likely cause of the reported event is due to improper handling by the user.Olympus will continue to monitor field performance for this device.
|
| |
|