Catalog Number CT318
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).A4: weight unknown / not provided b3: date of event unknown / not provided a summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost non-existent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site #16 was removed due to peri-implantitis.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 3331 and 4110.H10: narrative/data was updated.Dhr review was completed for the subject lot number 1266440.No deviations or non-conformance's, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record op# 120 ¿ str2 was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 1266440 for similar events and no other complaint was identified.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection, bone loss, and/or peri-implantitis problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EZTETIC DENTAL IMPLANT, 3.1MM DIAMETER, 2.9MMD PLATFORM, 8MM LENGTH
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key18839306
Report Number0002023141-2024-00591
Device Sequence Number7485385
Product Code DZE
UDI-Device Identifier00889024343917
UDI-Public(01)00889024343917(17)280423(10)1266440
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K142082
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberCT318
Device Lot Number1266440
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date06/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexMale
Patient EthnicityNon Hispanic
-
-