MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507470MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
02/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 60-year-old caucasian female patient underwent a primary breast augmentation with a 470cc mentor memorygel breast implant and experienced a rupture on the right side post-operatively, which was diagnosed during surgery.The patient underwent explantation on (b)(6) 2024, and replaced with a 790cc mentor memorygel xtra breast implant with serial number (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
On march 14, 2024, the mentor failure analysis lab has received the device for evaluation.On march 18, 2024, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned sample revealed that the sm mod classic gel, 470cc breast implant was found to be ruptured and received in four parts.In addition, a crease/fold was observed at the edges of the rupture.The evaluation determined that the possible cause of the rupture is consistent with normal wear.The instructions for use states that ruptures can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following may cause implants to rupture: stressing the implant during implantation or other procedures and weakening it; folding or wrinkling of the implant shell.Breast implants may also wear over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|