• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0701-NTW
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported this was a mitraclip to treat degenerative mitral regurgitation (mr) with a grade of 4.The steerable guide catheter (sgc) was inserted without issues.However, when the clip delivery system (cds) was inserted into the sgc hemostatic valve, air was observed in the sgc.This occurred multiple times; therefore, the sgc and cds were removed and replaced.It was noted the leak was more than likely caused by the cds.One clip was then successfully implanted, reducing mr to a grade of <1.There was no clinically significant delay in the procedure.
 
Additional Manufacturer Narrative
All available information was investigated, and the reported leaking clip introducer was confirmed via returned device analysis.Additionally, a torn clip introducer valve and missing silicone fluid were observed.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and the returned device analysis, the reported leak/splash was a cascading event of the observed torn clip introducer valve.The investigation determined the observed torn clip introducer valve and missing silicone fluid to be related to a potential product quality issue.The reported unexpected medical intervention was a result of case-specific circumstances.Per (b)(4), revision f, this complaint is within the scope of an exception escalation as the complaint description and device code match the specific issue described in the exception.Therefore, exception (issue) 126748 and exception (action) 131415 are referenced.The investigation evaluated the reported issue, and the engineering group determined the root cause to be man for the missing silicone and therefore, the torn clip introducer valve as the operators inadvertently missed the silicone verification.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key18839796
Report Number2135147-2024-01001
Device Sequence Number7532851
Product Code NKM
UDI-Device Identifier08717648230967
UDI-Public(01)08717648230967(17)240627(10)30629R1032
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2024
Device Catalogue NumberCDS0701-NTW
Device Lot Number30629R1032
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/13/2024
Supplement Date Received by Manufacturer03/27/2024
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date04/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/29/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STEERABLE GUIDE CATHETER
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-