A sales representative reported: "on (b)(6) 2022, the patient had hydrocephalus symptoms and a destroyed mechanism was seen on x-ray.They talked about it as part of a further training course." the valve was implanted on (b)(6) 2013; explanted and replaced on (b)(6) 2024.The boy has had a total of 7 mris in his life, the last one was on (b)(6) 2024.The valve defect also occurred during this time.
|
|
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10.The hakim valve was returned for evaluation: dhr - lot cpkb4k, conformed to the specifications when released to stock failure analysis - the valve was visually inspected; stator was dislodged.The valve was disassembled.The valve was visually inspected under microscope at appropriate magnification : biological debris was found on the motor and on the ruby ball.Traces of corrosion have been identified on stator.A visual control magnetizing engines was done, a normal polarization was noted.Root cause - -the root cause for the problem reported by the customer is due to the corrosion of the mechanism linked to the accumulation of biological debris which has weakened and destroyed the mechanism.The root cause for the programming issue is due to biological debris and protein build up interfering with the valve in view of the biological debris found during the investigation.
|