|
The site notified berlin heart inc.Clinical affairs that on (b)(6) 2024, a pump change was performed due to hemolyis and squeaking at the valves, louder at the outflow than the inflow valve.The patient's ldh level increased from 1853 on (b)(6) 2024 to 2494 on (b)(6) 2024.According to the site, the pump maintained full fill and ejection and functioned as intended.
|
|
The excor blood pump, s/n (b)(6), was in use on the patient until the pump exchange (14 days).On 2024-03-07 the blood pump in question was returned to berlin heart gmbh for investigation.We have reviewed the production records of the excor blood pump s/n (b)(6).This pump was produced according to our specifications.During initial visual inspection of the returned blood pump no abnormalities could be detected.The blood pump was tested for functional performance.Both the parameters from the clinic, as well as the parameters used for pump release during production have been taken into consideration.The pump performance met our specifications.The pump was completely filling and emptying.During the evaluation, a noise was detected emanating from the blood pump.In addition to the typical pump noise, a distinct "squeaking" sound could also be heard at the outflow valve of the blood pump.A slight vibration of the leaflets was detected.The analysis of the valve leaflets showed no defects.Squeaking noises were confirmed during evaluation.The valves were tested during production at berlin heart gmbh and no abnormalities were found.During operation on patients, a squeaking noise can sometimes occur from the pu valves.This effect only occurs in a valve-dependent pressure range in the pump valve junction.The pressure at the valve is strongly dependent on the patient's circular system, the implanted cannulas and the patient's blood properties.Therefore, the squeaking may start or stop when the drive parameters are changed or the patient's mobility changes.In this case, the reported squeaking noise at the valves could be reproduced during the analysis.A slight vibration of the valve leaflets was detected.Therefore, a correlation between the pump and the patient's elevated hemolysis cannot be excluded.Appropriate measures were implemented by the pump manufacturer, berlin heart gmbh.This pump was manufactured before the implementation of corrective action measures.
|