|
Pump was tested with the following protocol: connect tubing into reservoir.Prime tubing by gravity, then connect tubing to pressure gauge.Test 1: turn on the pump fail test if pump displays "pressure error" on post test 2: set prog to 125ml/hr for 20 vtbi, go to biomed options and set pressure to 8 psi.Run pump, pass pump if it alarms between 0 and 16 psi.Test 3: set prog to 125ml/hr for 20 vtbi, go to biomed options and set pressure to 30 psi.Run pump, pass pump if it alarms between 22 and 38 psi.Pump passed all 3 tests with no false or constant occlusion alarms.Pressure gauge s/n: (b)(6) cal date: on (b)(6) 2023.Reported issues is not confirmed.Pump meets passing criteria.Pump is being scrapped since it has reached end of life.
|