| Model Number |
E3520H |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Electric Shock (2554)
|
| Date of Event |
02/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, the machine was being used with a forcetriad.Nurse received an electric shock and burn at front of forcetriad in use with a force argon.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: forceargonii-8, forceargonii-8 force argon ii-8 x1 (serial#(b)(6).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: forcetriad, forcetriad force triad energy platform (serial# (b)(6)).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the machine was being used with a forcetriad.The nurse received an electric shock and burn at front of forcetriad in use with a force argon.When connecting e3520h lead to ufp (universal foot pedal) pencil activated by scrub nurse.The nurse had contact with exposed metal terminal on covidien vallylab argon handset e3520h.The burn was minor, and size and location of burn site was in thumb index finger.Nurse was checked in ed (emergency department) and kept under obs.
|
| |
|
Search Alerts/Recalls
|