| Model Number |
002003 |
| Medical Device Problem Code |
Partial Blockage (1065)
|
| Health Effect - Clinical Code |
Decreased Respiratory Rate (2485)
|
| Date of Event |
01/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Cannister has problems with the porous stone becoming blocked and the patient not getting flow.Patient not getting flow.
|
| |
|
Additional Manufacturer Narrative
|
|
There is an interference in therapy and ventilation.
|
| |
|
Event or Problem Description
|
|
Cannister has problems with the porous stone becoming blocked and the patient not getting flow.Patient not getting flow.
|
| |
|
Additional Manufacturer Narrative
|
|
There is an interference in therapy and ventilation.Based on the investigation and per pictures sent by the customer we can confirm the reported defect.Customer reported that the diffuser "porous stone" part number 65-4197 is blocked.Given that the reported component is purchased from a supplier, the supplier provides us with a certificate certifying that component 65-4197 meets the specified porosity according to our procedures.It's possible that the batch sent by our supplier of component 65-4197 did not have the required porosity for the production of (b)(4), leading to the reported defect."since the reported defect was confirmed the following actions were taken: a quality notification has been sent to our supplier of the component p/n 65-4197.An inspection will be conducted on the next 5 batches of part number 65-4197 arriving at our facility.".
|
| |
|