AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE Back to Search Results
Model Number CF-Q180AI
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the device was cleaned, sanitized, and disinfected prior to being sent to for repair.The device was returned, and the evaluation found no malfunctions aside from the allegation reported in b5.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was observed that during the device evaluation, the colonovideoscope exhibited a foreign object on the jet tube, the air/water cylinder, and the air/water tube.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.The foreign material could not be identified and the cause of the material remaining in the device could not be specified.There was no damage to the area where the foreign material was detected and it was confirmed that reprocessing was performed according to the instructions for use (ifu).The event can be detected and prevented by handling the device in accordance with the following ifu: evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 "preparation and inspection" shows how to detect the event.Evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 "reprocessing the endoscope (and related reprocessing accessories)" shows how to prevent the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18843202
Report Number9610595-2024-04654
Device Sequence Number20772776
Product Code FDF
UDI-Device Identifier04953170307690
UDI-Public04953170307690
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCF-Q180AI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer03/18/2024
Initial Report FDA Received Date03/05/2024
Supplement Report FDA Received Date03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-