Catalog Number AT75164
Medical Device Problem Codes Inflation Problem (1310); Unraveled Material (1664); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one atlas catheter was returned for evaluation.Distal tip of the balloon was noted to be inflated abnormally and the balloon was inflated with an unknown fluid.Unraveled fiber was also observed on the returned balloon.No other anomalies were noted during the visual evaluation.On functional testing, the balloon was deflated and again re-inflated to 6 atm and the balloon was still observed to be inflated in an abnormal condition.No other functional testing.One photo was reviewed.The photo shows the balloon in an inflated condition and the distal end of the balloon is bulged.The photo showed the balloon was not inflated uniformly along the length of the balloon.No other anomalies noted.As the submitted photo and returned sample analysis confirms the evidence of abnormal inflation and the balloon was noted to be bulged near the distal end of the balloon, the investigation is confirmed for the reported abnormal inflation.The investigation is also confirmed for the identified unraveled balloon fiber material as it was observed on the returned balloon material.A definitive root cause for the reported abnormal inflation and identified unraveled balloon fiber could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 03/2026) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly failed to inflate fully.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18844064
Report Number2020394-2024-00322
Device Sequence Number15527963
Product Code DQY
UDI-Device Identifier00801741062469
UDI-Public(01)00801741062469
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberAT75164
Device Lot NumberGFHQ1534
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2024
Initial Report FDA Received Date03/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-