| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported an isocenter mismatch message when sending the ct field for one patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H4 updated h6 updated h11 updated.The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer stated that they received an isocenter mismatch message when sending the ct field for one patient.It was confirmed through a review of the software log files, that the user was presented with a mismatched warning message and the intended isocenter values were used.Mosaiq did not have any malfunction and was working as designed and intended.Based on available information there was no mistreatment.Elekta was unable to reproduce the customer's noted issue where this potentially is a one time occurrence where the linear accelerator xvi imaging system delivered a mismatched value that was detectable.
|
| |
|
Search Alerts/Recalls
|