| Catalog Number |
PTA5-35-80-4-4.0 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown or unavailable.H3: device evaluated by mfg = other (81) - device evaluation has begun; however, a conclusion is not yet available.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
As reported, during an unknown procedure, an advance 35 lp low profile balloon catheter would not reach nominal pressure.Upon return and initial evaluation of the device, the balloon was ruptured.Additional information has been requested.
|
| |
|
Event or Problem Description
|
|
Additional information was received 14mar2024.The procedure involved an arteriovenous (av) dialysis shunt.The patient's anatomy was stenosed, and the left arm av anastomosis/radial-cephalic fistula was highly occluded.An unspecified 6-french sheath was used during the procedure, as well as an unspecified inflation device.The balloon was inflated two times to an unknown pressure.The balloon was not inflated within a stent, and blood was not noted in the inflation device.The balloon catheter was removed by itself, and a new balloon was used to successfully complete the procedure.A section of the device did not remain inside the patient¿s body.The patient was not hospitalized, and did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Blank fields on this form indicate the information is unknown, unavailable or unchanged.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary of event: as reported, during an unknown procedure, an advance 35 lp low profile balloon catheter would not reach nominal pressure.Upon return and initial evaluation of the device, the balloon was ruptured.Additional information was received 14mar2024.The procedure involved an arteriovenous (av) dialysis shunt.The patient's anatomy was stenosed, and the left arm av anastomosis/radial-cephalic fistula was highly occluded.An unspecified 6-french sheath was used during the procedure, as well as an unspecified inflation device.The balloon was inflated two times to an unknown pressure.The balloon was not inflated within a stent, and blood was not noted in the inflation device.The balloon catheter was removed by itself, and a new balloon was used to successfully complete the procedure.A section of the device did not remain inside the patient¿s body.The patient was not hospitalized and did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Investigation evaluation: reviews of the complaint history, device history record (dhr), instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection was conducted as well.The complaint device was returned to cook for investigation.A longitudinal tear was noted running the full length of the balloon.A document-based investigation evaluation was performed.A review of the device history record found no relevant non-conformances on the lot.A review of complaint history found no additional complaints for this lot number.The product ifu was reviewed, and the customer followed all relevant instructions related to this failure.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The information provided upon review of the dmr, dhr, ifu, returned device, and complaint file suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Based on the information provided and the results of the investigation, cook has concluded that the patient¿s anatomy likely contributed to this event, as the patient anatomy was stenosed, and the left arm av anastomosis/radial-cephalic fistula was highly occluded.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
| |
|